Correlating Rheological Measurements with Dynamic Oscillation

2022-07-02 08:33:12 By : Mr. Tom Xu

We use cookies to enhance your experience. By continuing to browse this site you agree to our use of cookies. More info.

The Thermo Scientific™ HAAKE™ MiniLab Series micro-compounder is well-known for being the ideal equipment, when it comes to compounding and processing tiny sample quantities of around 7 ml. The HAAKE MiniLab 3 is a small conical twin-screw extruder fitted with a bypass valve and a back flow channel, which allows users to regulate the residence time of the sample in the compounder.

Figure 1. The HAAKE MiniLab 3 micro-compounder.

The back flow channel’s patented design has two pressure transducers and a slit capillary flow channel, which are used to measure the pressure drop in the capillary. A shear stress can be calculated from the pressure drop and the geometry of the slit capillary. A shear rate is correlated from the designated screw speed and the measured back pressure. The shear stress and the shear rate values are used to calculate the relative sample viscosity at varied screw speeds. This article shows the correlation between the relative viscosity measurements completed on the new HAAKE MiniLab 3 micro-compounder, with results of absolute rheological measurements taken with a high end rheometer the Thermo Scientific™ HAAKE™ MARS™ 60 rotational rheometer.

For this study, two different LDPE grades (Lupolen 1800H and Lupolen 1800S from LyondellBasell) have been blended in different ratios. To have consistent sample preparation and a homogenous product, all samples have been pre-compounded on a Thermo ScientificTM Process 11 Twin-screw Extruder.

Table 1. Compounding of the samples.

a) HAAKE MiniLab 3 micro-compounder with pneumatic ram feeder

Set of co-rotating screws Software: Thermo Scientific™ HAAKE™ PolySoft OS Software for MiniLab N2 purge

b) HAAKE MARS 60 rotational rheometer

20 mm parallel plates measuring geometry Controlled test chamber (CTC) for temperature control N2 purge

Sample weight: 6.5 g Feeding speed: 50 rpm Testing temperature: 190 °C Inert gas: N2 purge Testing speed: Speed program from 50 rpm to 350 rpm (controlled via the HAAKE PolySoft OS Software)

Testing temperature: 190 °C Testing mode: Frequency sweep in controlled deformation (CD) mode Deformation: 1% Frequency range: 0.1 - 628 rad/s

After a preheating time of about 10 minutes, the pressure transducers of the MiniLab micro-compounder have to be calibrated at the chosen measuring temperature, to avoid any temperature effect on the pressure measurement. The extruder barrel and the pneumatic feeding piston are continually purged by a constant nitrogen flow to avoid the occurrence of oxygen during the tests, to stop degradation of the LDPE samples. The sample is then fed into the running extruder using the pneumatic feeding piston. After 1-2 minutes, the extrusion pressure is equilibrated, which signifies that the sample is correctly molten and ready for the rheological test. The rheological test itself is performed and regulated by the HAAKE PolySoft OS Software. The software launches the pre-programmed measuring steps shown in Figure 2, checks when steady state conditions are attained and then measures the pressure drop between the pressure transducers in the slit capillary channel to calculate the shear stress.

Figure 2. Pressure drops in slit capillary of HAAKE MiniLab 3 micro-compounder at different rotational speeds.

Figure 3 shows the results of two rheological tests with sample 1 (“LDPE1800 H”). Each measuring point in the flow curve corresponds to one screw speed on the HAAKE MiniLab micro-compounder. With growing shear rate the viscosity drops, because of the typical shear-thinning behavior of polymer melts. Furthermore, the viscosity curves of the two independent tests are almost identical, which shows the excellent repeatability of the test technique.

Figure 3. Apparent viscosity ηαππ and apparent shear stress ταππ as a function of the apparent shear rate for low density polyethylene. The results of two independent runs with the same polymer are presented.

The results of the rheological measurements conducted with all five compounded samples in one diagram can be seen in Figure 4. It is easy to see how the sample viscosity diminishes with the increase of LDPE1800S and the decreasing quantity of LDPE1800H in the compound.

Figure 4. Flow and viscosity curves for all compounds.

It is remarkable how the relative test results produced on the HAAKE MiniLab 3 correlate to the results performed on an absolute Rheometer. To verify this, one of the compounds (Sample 3 “LDPE1800 H1S1”) was analyzed in an oscillatory frequency sweep experiment on a HAAKE MARS 60 rheometer. For the comparison of the tests with the HAAKE MiniLab micro-compounder and HAAKE MARS rheometer the rule of Cox-Merz relation is applied. Empirically the two Scientists who gave the Cox-Merz relation their names discovered that the steady-shear viscosity measured as a function of shear rate could be directly compared to the dynamic complex viscosity measured as a function of angular velocity:

This relationship was found to be valid for many polymer melts and polymer solutions, but it rarely gives reasonable results for suspensions. The benefit of this Cox-Merz relation is that it is technically simpler to work with frequencies than with shear rates. Most of the time polymer melts cannot be calculated at shear rates lower than 50 1/s in a rotational rheometer in open sensor systems such as cone and plate, or parallel plate because of the elastic effects enountered. Therefore instead of measuring a flow curve in steady-state shear, one can more easily use the complex viscosity of dynamic testing1.

Figure 5 clearly illustrates how the relative viscosity measurement done on a HAAKE MiniLab micro-compounder compares with the absolute data acquired from a high end rotational rheometer. The experiment proves that measuring the change in flow behavior using pressure sensors in the slit capillary channel removes possible influences from screw forces and thereby supplies reliable rheological information. Measuring the rheological data directly during compounding offers a two-fold benefit to the Researcher. Firstly, it saves time when the measuring happens directly in the extruder and sample preparation can be neglected. But also structural variations within the sample that happen during compounding can be noticed directly and valuable process information can be provided on the spot. The HAAKE MiniLab 3 micro-compounder gathers the rheological data under process conditions. When a wider measuring range is needed, the HAAKE MARS rheometer is an ideal complementary extension of the experimental setup.

Figure 5. Comparison of viscosity data obtained from a measurement with a HAAKE MiniLab 3 extruder and a HAAKE MARS rheometer.

Figure 6. The HAAKE MiniLab 3 micro-compounder.

This information has been sourced, reviewed and adapted from materials provided by Thermo Fisher Scientific – Materials & Structural Analysis.

For more information on this source, please visit Thermo Fisher Scientific – Materials & Structural Analysis.

Please use one of the following formats to cite this article in your essay, paper or report:

Thermo Fisher Scientific – Materials & Structural Analysis. (2019, December 05). Correlating Rheological Measurements with Dynamic Oscillation. AZoM. Retrieved on July 02, 2022 from https://www.azom.com/article.aspx?ArticleID=14215.

Thermo Fisher Scientific – Materials & Structural Analysis. "Correlating Rheological Measurements with Dynamic Oscillation". AZoM. 02 July 2022. <https://www.azom.com/article.aspx?ArticleID=14215>.

Thermo Fisher Scientific – Materials & Structural Analysis. "Correlating Rheological Measurements with Dynamic Oscillation". AZoM. https://www.azom.com/article.aspx?ArticleID=14215. (accessed July 02, 2022).

Thermo Fisher Scientific – Materials & Structural Analysis. 2019. Correlating Rheological Measurements with Dynamic Oscillation. AZoM, viewed 02 July 2022, https://www.azom.com/article.aspx?ArticleID=14215.

Do you have a question you'd like to ask regarding this article?

AZoM speaks with Sona Dadhania, a Technology Analyst at IDTechEx, about the role 3D printing will play in the future of industrial manufacturing.

AZoM speaks with Ania Jolly, Head of Research and Business Engagement at the Henry Royce Institute, about advanced materials progress as well as industry-led initiatives toward sustainable manufacturing.

We spoke with Holiferm, a University of Manchester spin-off hoping to turn home and personal care products green with their eco-friendly biosurfactant products.

This product profile outlines the MAX-iR FTIR Gas Analyzer from Thermo Fisher Scientific.

Learn more about the JAM-5200EBM E-Beam Metal Additive Manufacturing System for 3D Printing.

This product profile outlines the features and benefits of the Contour X - 500 3D Optical Profilometer.

AZoM.com - An AZoNetwork Site

Owned and operated by AZoNetwork, © 2000-2022

p:top.scrollTo(duration=200)" class="scrollToTop">Top